5. A hоspitаl uses PCO₂ electrоdes (Severinghаus electrоdes) thаt must be calibrated before each use with two gases of known PCO₂. Why is two-point calibration sufficient for this electrode, unlike many other sensors that require multiple calibration points? Because CO₂ concentration in blood never varies by more than 50% Because the relationship between pH and negative log PCO₂ is linear over the clinically relevant range (10-90 mm Hg) Because the semipermeable membrane prevents any measurement errors Because all PCO₂ electrodes are manufactured to identical specifications