A pоrtаble аbdоminаl radiоgraph performed at a 100 cm SID typically requires a technique of 80 kVp and 20 mAs on the available piece of equipment. However, when entering the patient's room, you are only able to obtain a 70 cm SID. Which of the following mAs values will compensate for this change in SID?